PAGE 2-2
1097626, V
ER. 03
Use only approved SimplyGo accessories.
Repairs and adjustments must be performed by Philips Respironics authorized service personnel
only. Unauthorized service could cause injury, invalidate the warranty, or result in costly damage.
Periodically inspect electrical cords, cables, and the power supply for damage or signs of wear.
Discontinue use and replace if damaged.
To avoid electric shock, unplug the device and remove the batteries before cleaning the bag. DO
NOT immerse the device in any fluids.
Your home care provider is responsible for performing appropriate preventive maintenance at the
intervals recommended by Philips Respironics.
For proper operation, your device requires unobstructed ventilation. Always make sure any open-
ings in the case are not obstructed by items which may impede ventilation. Do not place the device
in a small closed space (such as a closet). The device should not be used adjacent to or stacked
with other equipment. For more information, contact your home care provider.
Do not use an extension cord.
Device operation above or outside of the voltage, breath rate, temperature, humidity and/or alti-
tude values specified may decrease oxygen concentration levels.
Never drop or insert any object into any opening.
Be aware that the electrical cord and/or tubing could present a tripping or strangulation hazard.
Use only power cords supplied by Philips Respironics for this device. Use of power cords not sup-
plied by Philips Respironics may cause overheating or damage to the device and may result in
increased emissions or decreased immunity of the equipment or system.
Do not operate without the battery installed and working. If primary power is lost with no battery in
place, the device will stop operating without warning the user. If the device must be operated with-
out the battery, the user should be aware that there is no backup power and no warning if primary
power is lost.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in this manual.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. See
the EMC section of this manual for distances to observe between RF Generators and the Sim-
plyGo device to avoid interference.
Place the device in a location so as to avoid pollutants and fumes