Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
SimplyGo and SimplyFlo are intended for use in the electromagnetic environment specified below. The
user of SimplyGo and SimplyFlo should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+
6 kV Contact
+
8 kV Air
+
6 kV Contact
+
8 kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
+
2 kV for Power
Supply Lines
+
1 kV for Input/Output
Lines
+
2 kV for Power
Supply Lines
+
1 kV for Input/Output
Lines
Mains power quality should be that of
a typical home or hospital
environment.
Surge
IEC 61000-4-5
+
1 kV Line to Line
+
2 kV Line to Ground
+
1 kV Line to Line
+
2 kV Line to Ground
Mains power quality should be that of
a typical home or hospital
environment.
Voltage Dips, Short
Interruptions and
Voltage Variations
on Power Supply
Input Lines
IEC 61000-4-11
<5% U
T
(>95% Dip in
U
T
) for 0.5 Cycle
40% U
T
(60% Dip in
U
T)
for 5 Cycles
70% U
T
(30% Dip in
U
T
) for 25 Cycles
<5% U
T
(>95% Dip in
U
T
) for 5 Seconds
< 5% U
T
(>95% Dip in
U
T
) for 0.5 Cycle
40% U
T
(60% Dip in
U
T)
for 5 Cycles
70% U
T
(30% Dip in
U
T
) for 25 Cycles
<5% U
T
(>95% Dip in
U
T
) for 5 Seconds
Mains power quality should be that of
a typical home or hospital
environment. If the user of the Device
required continued operation during
power mains interruptions, it is
recommended that the Device be
powered from an interruptible power
supply or battery.
Power Frequency
(50/60) Magnetic
Field
IEC61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical home or
hospital environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.