PAGE 7-51097626, VER. 03
7.2 REAR ENCLOSURE REPLACEMENT
Removal:
1. For SimplyGo, remove the Battery if inserted.
2. Remove the nine screws securing the Rear Enclosure to the Spine Assembly.
3. Adjust the device so the handle is facing you and the battery compartment is facing up.
4. Gently separate the Rear Enclosure from the front enclosure. For SimplyGo, you will expose the
battery wire harnesses.
5. For SimplyGo, disconnect the Battery Wire Harness from J1 and J3 on the Main PCA.
Installation: (for item identification, refer to removal instructions above)
1. For SimplyGo, connect the Battery Wire Harness to J1 and J3 on the Main PCA.
2. Carefully press the Rear Enclosure to the Front Enclosure. Ensure you do not kink any tubing or
pinch any wiring.
3. Secure the Rear Enclosure with the nine screws.