PAGE 7-131097626, VER. 03
7.6 EXHAUST VALVE/PRESSURE (FILL) VALVE REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. For Exhaust Valve (yellow dot) replacement, remove wire harness from location J9 or J8 on the
Main PCA and for Pressure (Fill) Valve replacement, remove wire harness from location J6 and J7
on the Main PCA.
3. Remove the four T10 Torx screws securing the valve(s) to the spine assembly.
NOTE
The Outer Valves are the Pressure (Fill) Valves and the inner valves are the Exhaust
Valves (Marked w/ Yellow dot). Proper orientation of the valve is required.