PAGE 7-151097626, VER. 03
7.7 PATIENT DELIVERY VALVE REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. Disconnect the Patient Delivery Valve from J16 on the Main PCA.
3. Remove the two screws securing the Patient Delivery Valve to the Check Valve Assembly.
Installation: (for item identification, refer to removal instructions above)
1. Ensure that the two O-rings are installed on the bottom of the Patient Delivery Valve and secure
the Patient Delivery Valve to the Check Valve Assembly using the two screws.
2. Connect the Patient Delivery Valve wire harness to J16 on the Main PCA.
3. Install the Rear Enclosure.