7.9 CHECK VALVE / GASKET REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. Remove the 6 T9 Torx screws.
3. Remove the Check Valve and Gaskets from the spine assembly.
CAUTION
Do not remove the two flat head screws that secure the Patient Delivery Valve.