Bubbled gas
Temp
dCO
2
(µM)
400 ppm CO
2
25 °C
13.6
5,000 ppm CO
2
25 °C
170
10,000 ppm CO
2
25 °C
340
15,000 ppm CO
2
25 °C
510
Tab. 7 CO
2
calibration example.
Correction table for CO
2
calibration
Correction table for the volume of calibration solution to be added into the vessel for the calibration step in respect of
maximum vessel volume.
Note: 1 ml of calibration solution added to the 400 ml cuvette with maximum volume of 465 ml results in final
concentration of 20 µM dCO
2
in the vessel.
FMT150 400 ml
FMT150 1000 ml
ml DIC free H
2
O in cuvette
ml calibration solution
ml DIC free H
2
O in cuvette
ml calibration solution
450
0.968
1,185
2.548
451
0.970
1,186
2.551
452
0.972
1,187
2.553
453
0.974
1,188
2.555
454
0.976
1,189
2.557
455
0.978
1,190
2.559
456
0.981
1,191
2.561
457
0.983
1,192
2.563
458
0.985
1,193
2.566
459
0.987
1,194
2.568
460
0.989
1,195
2.570
461
0.991
1,196
2.572
462
0.994
1,197
2.574
463
0.996
1,198
2.576
464
0.998
1,199
2.578
465
1.000
1,200
2.581
466
1.002
1,201
2.583
467
1.004
1,202
2.585
468
1.006
1,203
2.587
469
1.009
1,204
2.589
470
1.011
1,205
2.591
471
1.013
1,206
2.594
472
1.015
1,207
2.596
473
1.017
1,208
2.598
474
1.019
1,209
2.600
475
1.022
1,210
2.602
476
1.024
1,211
2.604
477
1.026
1,212
2.606
478
1.028
1,213
2.609
479
1.030
1,214
2.611
480
1.032
1,215
2.613
481
1.034
1,216
2.615
482
1.037
1,217
2.617
483
1.039
1,218
2.619
484
1.041
1,219
2.622
Tab. 8 Correction table for the volume of calibration.