Service
and
Maintenance
4
2.
Disconnect
the
defibrillator/monitor
from
the
ac
power
source.
battery
power
(the
BATT
CHRG
message
is
not
displayed).
Place
the
Standard
defibrillator
paddles
on
the
energy
meter.
Make
sure
the
defibrillator/monitor
power
is
on
and
it
is
operating
from
Warning
SHOCK
HAZARD.
When
charged
and
discharged
as
described
in
this
procedure,
the
LIFEPAK
9
defibrillator/monitor
discharges
approximately
360J
of
electrical
energy
through
the
defibrillator
paddles.
Unless
discharged
safely
as
described
in
this
procedure,
this
electrical
energy
may
Cause
injury
or
death.
Do
not
attempt
to
perform
this
procedure
unless
you
are
thoroughly
familiar
with
the
operation
of
the
LIFEPAK
9
defibrillator/monitor.
O
August
1993
Physio-Control
4.
Charge
the
defibrillator
to
360J
and
discharge
into
the
energy
meter.
Repeat
this
4
more
times
for
a
total
of
five
(5)
360J
discharges
within
2
minutes.
CONNECT
AC
POWER
message
appears,
the
Battery
Pack
has
failed
this
procedure.
Discard
the
Battery
Pack
and
discontinue
the
procedure.
If
the
instrument
continues
to
operate
after
step
4,
leave
the
defibrillator/monitor
operating
on
battery
power
for
45
minutes.
6.
At
the
end
of
the
45
minutes,
repeat
step
4.
If
the
instrument
shuts
down
due
to
power
loss
at
any
time,
the
Battery
Pack
has
failed
this
procedure.
Discard
the
Battery
Pack.
If
the
instrument
continues
to
operate
on
battery
power
after
repeating
step
4,
the
Battery
Pack
has
passed
the
procedure
and
has
useful
life
remaining.
(It
is
acceptable
for
the
LOW
BATTERY:
CONNECT
AC
POWER
message
to
appear
and
remain
on
at
this
point.)
Recharge
the
Battery
Pack
for
24
hours
before
placing
into
service
or
storage.
If
the
instrument
shuts
down
due
to
power
loss
or
if
the
LOW
BATTERY:
Note:
Physio-Control
Corporation
recommends
replacing
the
Battery
Pack
installed
in
the
LIFEPAK
9
defibrillator/monitor
after
two
years
of
service
as
a
part
of
routine
maintenance.