Service
and
Maintenance
4
Power
Supply
Complete
Case
Separation.
Refer
to
Figure
4-2,
page
4-9:
Removal
1.
Disconnect
J1
on
the
Power
Supply
PCB
(A2).
2.
Remove
four
screws
(41)
from
the
rear
panel
(31)
and
remove
the
rear
panel.
3.
Disconnect
J2
on
the
Power
Supply.
4.
Remove
four
screws
(41)
from
the
Power
Supply
and
detach
from
chassis
(7).
Interconnect
PCB
Complete
Case
Separation.
Refer
to
Figure
4-2,
page
4-9:
Removal
1.
Remove
the
Power
Conversion
PCB,
the
Main
PCB,
the
Display
PCB/CRT
Assembly,
and
the
Power
Supply
PCB.
2.
Disconnect
J29.
3.
Remove
four
screws
(41)
from
the
Interconnect
PCB.
Inspection
When
servicing
or
repairing
the
instrument,
routinely
begin
with
a
visual
Techniques
inspection
of
the
hardware
and
components
for
signs
of
damage.
Signs
of
damage
require
further
inspection
of
the
surrounding
area
for
peripheral
breakage
or
damage.
Exterior
Inspection
Visually
inspect
the
entire
instrument
for
wear,
corrosion,
deterioration,
and
damage
resulting
from
extreme
temperatures
or
dropping.
Lift
and
hold
the
instrument
upside
down
while
listening
for
loose
hardware.
Interior
Inspection
Table
4-1,
Inspection
Techniques,
lists
the
major
hardware
components
of
the
instrument,
possible
problems
associated
with
each
item,
and
recommended
corrective
actions.
O
August
1993
Physio-Control
All
Rights
Reserved
4-11