Testing
3
3.
Disconnect
one
lead
of
the
defibrillation
cable
while
the
defibrillator
is
still
analyzing.
Three
beeps
will
sound
and
the
following
CRT
message
will
flash:
¢
TO
ANALYZE:
CONNECT
DEFIB
PADS
4.
Reconnect
the
lead
to
the
Patient
Simlator
and
three
beeps
will
sound
and
the
CRT
will
display;
¢
ANALYZING
NOW,
STEP
BACK
PIP-Motion
Detection
1.
Press
VF
on
the
patient
simulator,
then press
and
hold
the
MOTION
pushbutton
to
send
a
simulated
motion
signal
into
the
instrument.
2.
Press
ANALYZE
and
verify:
e
The
CRT
displays
TO
ANALYZE:
STOP
MOTION
e
A
priority
three
tone
sounds
3.
Continue
to
hold
down
the
MOTION
pushbutton
an
additional
30
seconds
and
verify:
e
The
CRT
displays
MOTION
DETECTED
CANNOT
ANALYZE
e
A
priority
three
tone
sounds
4.
After
another
20
seconds,
verify
the
ANALYZE
LED
is
off
and
the
motion
message
disappears.
PIP-Test
Load
1.
Connect
the
defibrillator
cable
to
the
test
load
contacts
and
check
that
a
roll
of
paper
is
in
the
recorder.
2.
Press
CHARGE
on
the
adapter.
3.
Press
both
discharge
buttons
simultaneously
on
the
adapter
and
verify:
e
The
recorder
annotates
the
appropriate
data.
e
The
CRT
displays
TEST
200
JOULES
DELIVERED.
©
November
1994
Physio-Control
Corporation
All
Rights
Reserved
3-5