LIFEPAK
9
shock
advisory
adapter
Service
Manual
3-6
PIP-Leakage
Current
Perform
an
ac
leakage
check
using
a
safety
analyzer
with
the
adapter
installed
in
the
LIFEPAK
9
defibrillator/monitor.
Check
leakage
current
in
accordance
with
the
following
standards:
AAMI/ANSI
(Association
for
the
Advancement
of
Medical
Instrumentation/American
National
Standards
Institute)
ES1-1985,
AAMI/ANSI
DF2-1989,
CSA
(Canadian
Standards
Association)
C22.2,
IEC
(International
Electrotechnical
Commission)
601-1,
IEC
601-2-4,
and
UL
(Underwriters
Laboratories)
544.
1.
Lead
Source
Current
(Patient
Cable):
201A
or
less
2.
Lead
Sink
Current
(Patient
Cable):
20LA
or
less
3.
Defibrillation
Cable
sink
current:
100pA
or
less
(typically
551A)
©