EasyManua.ls Logo

Physio Control LIFEPAK 9 - Page 36

Physio Control LIFEPAK 9
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LIFEPAK
9
defibrillator/monitor
This
preserves
the
dc
offset
level
required
by
the
Echo
circuit.
DC
Restore
controls
any
transient
signals
to
prevent
the
high
potential
gain
of
the
Instrumentation
Amplifier
from
falsely
amplifying
the
dc
level
input
for
conversion.
Under
normal
operation,
the
A/D
conversion
dc
level
at
TP19
is
from
+1V
to
+11V.
Switch
smoothing
is
the
process
of
limiting
the
response
of
the
preamp
to
dc
offset
voltages
when
switching
from
one
lead
(I,
Il,
Ill,
or
STD)
to
another.
This
function
is
initiated
when
the
appropriate
number
of
pulses
(see
Table
1-3,
pulses
12
thru
20) turns
on
Q8
or
Q9
and
switches
U42
to
the
selected
lead.
When
Q8
switches
on,
C90
discharges
from
+12V,
causing
Q11
to
conduct.
Q11
changes
the
high-pass
filter
by
switching
R161
in
parallel
with
R165.
Because
of
R161,
this
filter
limits
the
signal
passband
to
above
40Hz.
Transient
signals
associated
with
defibrillation
are
sensed
by
op
amp
U45C.
U45C
functions
as
a
window
comparator
and
responds
with
a
positive
output
when
the
voltage
at
its
input
exceeds
+2V.
Resistors
R172
and
R147
apply
a
negative
bias
to
the
noninverting
input
of
U45C;
R171
and
R143
apply
a
positive
bias
to
the
inverting
input.
C104
and
R151
function
as
a
low-pass
filter
to
prevent
very
short
spikes
(such
an
an
internal
pacing
spike)
from
inducing
DC
Restore.
The
output
of
U45C
drives
transistor
Q10,
routing
the
signal
through
high-pass
filter
U44C
and
U45D.
Resistor
R160,
with
U44C
and
U45D,
limits
the
signal
passband
to
5Hz.
When
PADDLES
is
selected,
further
passband
limiting
is
required
to
eliminate
motion
artifact
introduced
by
the
paddles
when
used
in
the
QUIK-LOOK
mode.
A
command
pulse
selects
paddles
mode,
(see
Table
1-3,
pulse
20),
turning
Q9
on.
When
Q9
conducts,
flip-flop
U36B
is
loaded
so
that
the
next
clock
pulse
at
pin
6
switches
the
Q
output
high.
Analog
switch
U43B
then
closes,
bringing
R155
in
parallel
with
R165
in
the
high-pass
filter,
and
limiting
the
ECG
signal
passband
to
above
2Hz.
Echo.
The
ECG
signal
is
converted
to
a
pulse-delay
period
on
command
from
the
System
microprocessor.
The
time
from
when
the
convert
command
from
the
System
microprocessor
is
received
until
the
“echo”
pulse
is
sent
back
to
the
System
microprocessor
is
proportional
to
a
sample
of
the
ECG
signal
(see
Figure
1-6).
When
the
echo
pulse
is
sent
back
to
the
System
microprocessor
(across
an
opto-isolator),
the
System
microprocessor
converts
the
pulse-delay
period
to
a
number.
1-22
All
Rights
Reserved
©
August
1993
Physio-Control

Table of Contents

Other manuals for Physio Control LIFEPAK 9

Related product manuals