Testing
3
TCP
Checklist
Customer:
Date:
Dept/Location:
Instrument
Type:
Model:
Serial
Number:
Frequency:
Last
Insp:
Tech.
Name/ID:
Test
and
Calibration
Pass
Fail
Comments
Voltage
Checks
Low
Battery
Threshold
Checks
Monitoring
Check
Defib
Charge
Check
Test
Menu
Access
Defibrillator
Calibration
O
Selected
Acceptable
Range
MEASURED
|
Selected
Acceptable
Range
MEASURED
Value
(4)
value
(4)
%
Tolerance
VALUE
|
Value
(5)
Value
(4)
%Tolerance
VALUE
(J)
1
0.8
to
1.2
+20%
20
19.0
to
21.0
15%
2
1.8
to
2.2
+10%
30
28.5
to
31.5
+5%
3
2.7
to
3.3
+10%
50
47.5
to
52.5
+5%
4
3.8
to
4.2
+5%
100
95.0
to
105.0
+5%
5
4.8
to
5.2
+5%
200
190.0
to
210.0
+5%
6
5.7
to
6.3
+5%
300
285.0
to
315.0
+5%
7
6.7
to
7.3
+5%
360
342.0
to
378.0
+5%
8
7.6
to
8.4
+5%
9
8.6
to
9.4
+5%
10
9.5
to
10.5
+5%
Test
Load
Calibration
Charge
Time
Output
Waveform
O O
Display
Calibration
О
Power
Supply
Calibration
Recorder
Calibration
Recorder
Frequency
Response
CRT
Frequency
Response
ECG
Output
170008-001
Orig
©
August
1993
Physio-Control
All
Rights
Reserved
3-39