Page 43 of 43
Guidance and manufacturer´s declaration – electromagnetic emissions/
immunity
The plusoptiX A09 is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is in such an environment.
Electromagnetic environment - guidance
The plusoptiX A09 uses RF Energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The plusoptiX A09 system is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low
Voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC
61000-3-3
Electromagnetic environment –
Guidance
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material the relative humidity should be at
least 30 %
Electric fast transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 2 kV for power supply
lines
Mains power supply should be that of a
typical commercial or hospital environment
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power supply should be that of a
typical commercial or hospital environment
Voltage Dips, short interruptions and
voltage variations on power supply
input lines IEC 61000-4-11
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
Mains power supply should be that of a
typical commercial or hospital environment
Power Frequency (50/60 Hz) magnetic
field IEC 61000-4-8
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.