35
5.4.4. Loop
The option is available if the number of segments is equal to 2 or more.
When the program finishes the last segment, the device starts the program again from the first segment.
The selected number determines the amount of the program repetitions. If the time of the last segment is set
to infinity, it will be treated as infinite in the last cycle. In the remaining cycles it will be treated as 0.
5.4.5. Priority
Can be set in terms of:
Parameters:
The program without a ramp – the device starts the countdown of the segment time when the set temperature
is reached.
The program with a ramp – first, the device counts down the time of the ramp and then proceeds to the segment
countdown when the set temperature is reached. Regardless of whether the time of ramp elapsed.
Time:
The program without a ramp – the device starts counting down the segment time when the program is started.
Regardless of whether the temperature has been reached.
The program with a ramp – the device first counts down the ramp time and after its expiry it proceeds to the
countdown of the segment time. Regardless of whether the temperature has been reached.
It may happen that the device couldn’t reach the set temperature within the set time because
the reaching time was too short. In such situation the reaching time will be exceeded and the
segment’s time countdown will start when the set temperature is reached.
It may happen that the time of reaching was too short and the device failed to reach the set
temperature within the set time. Then the countdown of the segment time will start before reach-
ing the set temperature. Thus, the actual time of device stay in the set temperature will be
shortened.