10
3. BEFORE THE FIRST USE
By default, the unit is sent in a cardboard box. It is necessary to transport it in the upright position and
prevent it from any unintended movements.
The surface components of the unit are made of stainless steel, therefore slight discoloration may occur. It is
a result of the technologies used in the production of metal sheet in accordance with the requirements of EN
10088-2 standard and it is not a defect of the unit.
3.1. Installation place
The unit should be placed at a minimum distance of 100mm from the wall of the room. The height of the room
must be higher than the height of the unit by min 300mm.
In drying ovens (SL) and in incubators (CL) on the back of the unit, there is an air-flap located, through which
the hot air is extracted. The manufacturer recommends using a non-flammable insulation screen on the wall
or increasing the distance from the wall. Failure to do so may result in permanent damage to the wall and in
extreme cases, to a fire.
The place of location of the unit should meet the following conditions:
• Ambient temperature between +10°C...+26°C *
• Ambient relative humidity of 60% *
• Low dust environment
• Appropriately ventilated according to its size
• Placed on a hard and stable surface
• Placed at least 100 mm away from the wall
• Height of the room must be at least 300 mm greater than the height of the unit
• Keep away from direct sunlight
• The unit should be kept far from heat sources, air conditioning vents and drafts
• Kept away from any heat sources *
• Not designed to be built-in e.g. within a wall cavity
• The place of installation of the unit should contain a mains socket 230V/50Hz or 400V/50Hz
(SLW/SRWP 400/750/1450).
Failure to do so may result damage to the unit and losing the warranty rights.
Once you receive the unit, please check its the technical condition and all accessories.
Any claims regarding latent defects should be reported to the manufacturer, while any
damage during transport or incomplete accessories need to be passed to the entities
who are responsible for the transport and unloading.
Only for ILW, ST, CHL: While carrying the unit, please do not tilt it to one side more than
45° from the upright position, as there is a high probability of the damaging the com-
pressor. If it is necessary to tilt it to one side more than 45°, then after placing it, please
wait at least 2 hours before connecting the unit to the mains.