8. TEMPERATURE PROTECTION
Temperature protection is included as standard, if damage occurs to the temperature controller or the user
changes the temperature settings outside of the upper and lower limits, temperature protection will activate.
The figure below shows how this works.
Protection Class 3.3. according to DIN 12880 norm is called protecting sample function. The User can set the
temperature protection lower / upper by himself. When the set temperature falls outside of this range, the
power of cooling system or heating system is switched off.