A-2
A TMS-A-305 Calibration Set
A.2 Defining the Depth Setting Standard
Before you can make a measurement using the depth setting standard, you
first of all have to define it, i.e. apply the data from the calibration certificate to
the file Standards.xml. If you want to use other standards, you also have to
define them first.
INFORMATION
If you define just one standard, you will not have to select a standard later
during calibration.
INFORMATION
To define depth setting standards, you require write permissions in the
directory C:\Program Files\Polytec.
To define the depth setting standard, you proceed as follows:
1. Open the file Standards.xml in a text editor. You will find the file in the
following directory:
C:\Program Files\Polytec\TMS
X.X\Pipeline\AddIns\CalibrationDataEvaluation