DefiMonitor XD 21324 / GB / K05
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Internal battery voltage too high
- restart the device
- if the error persists: Return the
device for servicing
Operating temperature too high
- operate the device in a cooler
environment
Operating temperature too low
- operate the device in a warmer
environment
Error communicating with power
module
- restart the device
- if the error persists: Return the
device for servicing
Error during the shock appliance
- return the device for servicing
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
Power module voltage too low
- charge and/or replace the
power module
Internal communication error
- restart the device
- if the error persists: Return the
device for servicing
Previous error detected -
need full test
General description of error
detected during auto-diagnosis
- restart the device
- if the error persists: Return the
device for servicing
Program code area
corrupt
Error in the program memory
- return the device for servicing