List of error codes
84 / 116
DefiMonitor XD 21324 / GB / K05
device for servicing
104
24V exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
105
24V drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
106
PWRIN exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
107
PWRIN drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
108
Battery exceed
Internal battery voltage too high
- restart the device
- if the error persists: Return the
device for servicing
110
Temperature exceed
Operating temperature too high
- operate the device in a cooler
environment
111
Temperature drop
Operating temperature too low
- operate the device in a warmer
environment
116
HDQ interface fail
Error communicating with power
module
- restart the device
- if the error persists: Return the
device for servicing
117
CYCLE time error
Error during the shock appliance
- return the device for servicing
118
VREF exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
119
VREF drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
120
Battery/accumulator fail
Power module voltage too low
- charge and/or replace the
power module
121
DSP timeout error
Internal communication error
- restart the device
- if the error persists: Return the
device for servicing
122
Previous error detected -
need full test
General description of error
detected during auto-diagnosis
- restart the device
- if the error persists: Return the
device for servicing
123
Program code area
corrupt
Error in the program memory
- return the device for servicing