DefiMonitor XD 21324 / GB / K05
Error in the membrane keyboard
- restart the device
- if the error persists: Return the
device for servicing
Error during the ECG calibration
- restart the device
- if the error persists: Return the
device for servicing
Clocks (32 kHz or
3.6864 MHz) error
Error in system clock generation
- restart the device
- if the error persists: Return the
device for servicing
Error during the hardware
detection
- return the device for servicing
Internal communication error
- restart the device
- if the error persists: Return the
device for servicing
Error when switching the
paddles
- detach completely the device
from the current supply:
Disconnect the power plug,
remove the power module from
the device; Restart the device
after an average time of 30
seconds
- if the error persists: Return the
device for servicing
XtBoard err - power
source
Error during the current supply
of the expansion board
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
XtBoard err - paddle logic
Error during the test of the
paddles
- do not press any paddle
buttons when starting the device
- remove and reinsert the
paddles
- restart the device
- if the error persists: Return the
device for servicing
Error during the test of the
paddle buttons
- restart the device
- do not press any paddle
buttons when starting the device
- if the error persists: Return the
device for servicing
General fault message
expansion board
- restart the device
- if the error persists: Return the
device for servicing
Defective storage of important
variables
- return the device for servicing
XtBoard err - runtime error
Error during the expansion
board runtime check
- restart the device
- if the error persists: Return the
device for servicing
Error during the impedance
- return the device for servicing