88 PRODIGIT
[LIMit
]POWer
{HIGHLOW} or WHWL
Syntax [LIMit]POWer{ HIGHLOW}{SP}{ NR2 }{NL}
[LIMit]POWer{ HIGHLOW} {NL}
[WHWL]{SP}{ NR2 }{NL}
[WHWL]{NL}
PurposeTo set the upper/lower limit value of threshold power (W).
DescriptionThis command is to set the upper/lower limit value of threshold power
(WATT). When power (WATT) is lower than this lower limit value or
higher than the upper limit value, NG indicating light will come on to
indicateNO GOOD.
[LIMit
] VOLtage
{HIGHLOW} or VHVL
Syntax [LIMit] VOLtage{ HIGHLOW}{SP}{ NR2 }{NL}
[LIMit] VOLtage{ HIGHLOW} {NL}
[VHVL]{SP}{ NR2 }{NL}
[VHVL}{NL}
PurposeTo set the upper/lower limit value of threshold voltage.
DescriptionThis command is to set the upper/lower limit value of threshold voltage.
When input voltage is lower than the lower limit value or higher than the
upper limit value, NG indicating light will come on to indicateNO GOOD
.
[LIMit
] SVHSVL
Syntax [LIMit] {SVHSVL}{SP}{ NR2 }{NL}
[LIMit] { SVHSVL} {NL}
PurposeTo set the upper/lower limit value of short current.
DescriptionThis command is to set the upper/lower limit value of short current. When
short current is lower than the lower limit value or higher than the upper
limit value, NG indicating light will come on to indicateNO GOOD.