EasyManua.ls Logo

Protec OPTIMAX 2010 NDT - Maintenance

Protec OPTIMAX 2010 NDT
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPTIMAX
2010 NDT
User Manual
516
0
-
0
-
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
33
von
58
2.22 Activate regeneration pumps for at least five minutes
2.23 Check all pumps for tightness
2.24 Switch machine off
2.25 Drain tanks
2.26 Fill tanks 3/4 with water
2.27 Switch machine on
2.28 Activate replenishment pumps manually until tanks overflow
2.29 Activate film transport for a few minutes
2.30 Switch machine off and drain all three tanks
2.31 Repeat item 2.20 to 2.30 if required (observe information e. g. concerning temperature, time
and cleaning procedure outlined in tank cleaner datasheet)
2.32 Remove roller racks from the machine and remove dirt under flowing water using a soft rag or
sponge
2.33 Clean all toothed gear wheels, axles, bearings and rollers, check them for damage (replace if
required)
2.34 Remove light protection flap and wipe its underside using a soft rag
2.35 Clean inlet plate using a soft rag
2.36 Mount the cover sheet of the film intake
2.37 Align roller racks and re-insert them in machine
2.38 Fill machine with chemicals
2.39 Switch machine on
2.40 Adjust bath temperature to previously adjusted value
2.41 Feed cleaning film (approx. 4 pieces)
2.42 Check function as described under item 1.
2.43 Approx. 15 minutes after reaching of the bath temperature, check it by measuring and
recalibrate if required (see 7.4.5)
3 Perform constancy tests based on the applicable national standards (e.g. IEC61223-2-1 and DIN
6868-2).
6.3.2 Maintenance
Required maintenance must be performed at 3-6 month intervals, depending on the film throughput,
by PROTEC Service or specific authorized service provider to ensure the sage and reliable operation of
the equipment.
In the event that scheduled maintenance is not performed, PROTEC GmbH & Co. KG will not be
responsible for damages incurred by the user or third parties if such damages are the result of improper
or omitted maintenance.
Prior to operation, the operator must ensure that all Safety related mechanisms, indicators and/or
switches described within the user manual are fully functional and that the unit is overall operationally
ready.
See Technical Description off the system and off all integral components.
Only original spare parts are to be used in situations requiring component replacement.

Table of Contents