What to do if my Pulsar Controller displays an incorrect reading for the current level?
- HHannah RamosJul 27, 2025
If the Pulsar Controller displays an incorrect reading, measure the actual distance from the transducer head to the surface of the material. Then, enter Program Mode and directly access P21 (Set Distance). Type in the measured distance, press ENTER, and then ENTER again when prompted. Wait until SET is displayed and return to Run Mode. The display should now update to the correct reading.