ZENITH 140 INSTRUCTION MANUAL
132
Bargraph
P829 Bargraph
By default, the bar graph will be representative of the reading obtained, as
determined by the Mode P100. When P100 = 4 (Average) or 5
(Differential) the bar graph can be assigned to be representative of the
level on any of the two points of measurement. This parameter is
automatically set to the correct default option when selecting the Mode
P100 and Xducer (P101), and under normal circumstances will not require
changing. The options, dependant on the value entered for Mode P100,
are as follows:
OPTION
DESCRIPTION
1 = Auxiliary (Optional)
Bargraph will be representative of levels obtained
from optional Aux. Input.
2 = Xducer 1 (Default)
Bargraph will be representative of levels obtained
from Xducer 1
3 = Xducer 2
Bargraph will be representative of levels obtained
from Xducer 2
4 = Avg. Level or Diff
Bargraph will be representative of levels obtained
from optional Xducer 1