EasyManuals Logo

Quantech QTC3040T User Manual

Quantech QTC3040T
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
Quantech
187
Section 9: Service and troubleshooting
Form QTC3-NM1
Issue date: 12/19/2023
9
Table 33: Pressure Transducers
0-400 PSIG SUCTION
PRESSURE
TRANSDUCER
0-600 PSIG DISCHARGE
PRESSURE
TRANSDUCER
PRESSURE
PSIG
VOLTAGE
VDC
PRESSURE
PSIG
VOLTAGE
VDC
0 0.5 0 0.5
50 1.0 75 1.0
100 1.5 150 1.5
150 2.0 225 2.0
200 2.5 300 2.5
250 3.0 375 3.0
300 3.5 450 3.5
350 4.0 525 4.0
400 4.5 600 4.5
RED WIRE = 5V, BLACK WIRE = 0V, WHITE/GREEN WIRE = SIGNAL
TEST POINTS:
Suction Pressure:
System 1:
...............................................
Microboard J7-10 to J7-9
System 2:
...............................................
Microboard J9-10 to J9-9
Discharge Pressure:
System 1:
...............................................
Microboard J7-11 to J7-7
System 2:
...............................................
Microboard J9-11 to J9-7
V = (Pressure in psig x .01) + 0.5
or
V = (Pressure in barg x .145) + 0.5
where V = DC voltage output
Pressure = pressure sensed by transducer
The I/O board connections for the Discharge
Transducers are as follows.
System 1 discharge transducer
J7-6 = +5 VDC regulated supply to transducer.
J7-11 = VDC input signal to the microboard. See the
formula above for voltage readings that cor-
respond to specific discharge pressures.
J7-7 = +5 VDC return
J7-2 = drain (shield connection = 0VDC)
System 2 discharge transducer
J9-6 = +5 VDC regulated supply to transducer.
J9-11 = VDC input signal to the microboard. See the
formula above for voltage readings that correspond to
specific discharge pressures.
J9-7 = +5 VDC return
J9-2 = drain (shield connection = 0 VDC)
The suction transducers have a range from 0 to 400
psig (27.5 barg). The output will be linear from
0.5 VDC to 4.5 VDC over the 400 psig (27.5 barg)
range. Following is a formula that can be used to verify
the voltage output of the transducer. All voltage read-
ing are in reference to ground (unit case).
V = (Pressure in psig x 0.02) + 0.5
or
V = (Pressure in barg x 0.29) + 0.5
where V = DC voltage input to microprocessor
Pressure = pressure sensed by transducer
Following are the I/O board connections for the Suc-
tion Transducer.
System 1 suction transducer
J7-5 = +5 VDC regulated supply to transducer.
J7-10 = VDC input signal to the microboard. See the
formula above for voltage readings that cor-
respond to specific suction pressures.
J7-9 = +5 VDC return.
J7-1 = drain (shield connection = 0 VDC).
System 2 suction transducer
J9-5 = +5 VDC regulated supply to transducer.
J9-10 = VDC input signal to the microboard. See the
formula above for voltage readings that
correspond to specific suction pressures.
J7-9 = +5 VDC return.
J7-11 = drain (shield connection = 0VDC).
Digital outputs
See the unit wiring diagram and Figure 52 on page
188. The digital outputs are located on TB7, TB8, and
TB9 and TB-10 of the microboard. All outputs are 120
VAC with the exception of TB8-6 to TB8-7 which are
the contacts that can be used for a remote evaporator
pump start signal. The voltage applied to either of these
terminals would be determined by field wiring.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantech QTC3040T and is the answer not in the manual?

Quantech QTC3040T Specifications

General IconGeneral
BrandQuantech
ModelQTC3040T
CategoryChiller
LanguageEnglish

Related product manuals