EasyManua.ls Logo

Quidel Triage MeterPro - Page 35

Quidel Triage MeterPro
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page35of86
USERACTIVE(USRAONSCREEN)Atthetimeofpatienttesting,theuserhastheabilityto
deselectthetest.
6. Pressthekeytosavechanges.
QCTestsWithTestSettingsActivated
AnytestwhichtheSupervisorhassettoACTIVE,USERACTIVEorUSERINACTIVEwillrunontheQCSample
panel.Themeterwilldisplayandprintthetest’sQCsampleresultandupdatetheQCSampletimerforthe
test.IftheSupervisorhassetthetesttoINACTIVE,
thetestwillhavetheQCSampleturnedOFF.TheQC
Sampleresultwon’tshowforthattestandtheQCSampletimerwon’tbeupdatedforit.
IftheSupervisorchangesthetestfromINACTIVEtoACTIVE,toUSERACTIVEortoUSERINACTIVEthattest
maybeout
ofQCSampledate.Asalwaysforatestwithanexpir edQCSample,apassingQCSampletestwill
needtorunbeforeapatientsamplecanreportaresultforthattest.
QCParameters
TheQCParametersmenuprovidestheusertheoptionsfordeterminingwhatthe
maximumfrequencyisforrunningtheQCDeviceandQCSample,whether1or2
levelsofcontrolarerequiredfortheQCSample,andwhethertheallowedQC
SampleRangeofquantitativetestsis2or3
standarddeviations.
QCDeviceFrequency
TheQCDEVICEFREQUENCYsettingdeterminesthemaximumintervalbetweenQCDeviceteststhatauser
isallowedtorunpatienttests.Whentheintervalhaslapsed,allusersarelockedoutoftheRunTestmenu
untilaQCDevicehasbeensuccessfullyrun.
ItisrecommendedthattheQCDeviceberundailywhenperformingpatienttesting.
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERSusingthe
keys.
2. Pressthekey.
3. Usethe
keystoselecttheinformationtotherightofQCDEVICEFREQ.
4. Usingthe
keys,selectoneoftheseoptions:NONE,8HOURS,DAILY,WEEKLY,MONTHLY.
5. Pressthekeytosavechanges.
QCSampleFrequency
TheQCSAMPLEFREQUENCYsettingdeterminesthemaximumintervalbetweenQCSampleteststhatauser
isallowedtorunpatienttestsonapar ticulardevicelotnumber.Whentheintervalhaslapsed,allusersare
lockedoutoftheRUNPAT IENT TESTmenuuntilaQC
Samplehasbeensuccess fullyrun.Toviewwhena
particulardevicelot’sQCSamplewillexpire,usetheRECALLREAGENTLOTSQCfeature.
ItisrecommendedthattheQCSampleberunwitheverynewshipment,newlotormonthly,whicheveris
lesswhenperformingpatient testing.
NOTE:
Whentheanalyteisdeselected,thevalueis notcalculatednorstored
MainMenu
SetParameters
QCParameters