Page43of86
7. Presstheke y.
8. EntertheQClotnumberfromthelabelonthesideofthebottlevialcontaining theQCSample.
9. Tocorrect,pressthe keytocleartheentirenumber.
Thentypeinthecorrectnumber.
10.Pressthekey.
11.Performthetestaccordingtotheinstructionsprovided
inthepackageoftestdevicesyouareusing.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeterPrountilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c.
TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION:Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
12.Repeatstepsforeachqualitycontrolsample.
IMPORTANT:ImportantinstructionsforQCSampleTests
D Ifthetestdevice’sinternalqualitycontrol(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
b. IfanyoftheQCSampleresultsareoutofrange,theresultsforthatparticularanalytewill
bein
reversevideo(lighttextonadarkbackgr ound).
c. Apanelwithmultipleanalyteswillstillbeabletoreportpatientresultsonthoseanalyteswhich
passedQC.AnanalytethatfailedQCwillnotbereportedonpatienttests(insteadwillbeflagged
witha#symbol).
NOTE:OnlyenterthefourͲdigitnumericvalueoftheQClotnumber.Donotentertheprecedingalpha
character.
NOTE:IfyouenteraQClotnumberforwhichthereisnodataintheQuidelTriageMeterPro’smemory,you
willseeamessageonthemeter’sscreen:NOQCSampledatainmemory.
Tocorrectthis:InstalltheQCSampleCODECHIP(seepictureonpage37).
Press
ENTERtocontinue.