Misc.TestSample
1. PressthekeytopowertheQuidelTriageMeterProon.
2. SelectRUNTESTusingthe
keys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODE CHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
3. Pressthekey.
4. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
5. Pressthekey.
6. SelectMISC.TEST
usingthe
keys.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISCTEST
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
7. Pressthekey.
8. TypeorscantheMisc.Testidentification(Misc.TestID).Tocorrect,presstocleartheentireID,or
selecttheincorrectcharacterusingthe
keys.ThentypeinthecorrectID.
9. Pressthekey.
10.Iftheinformationis correct,pressthekey
toconfirmtheMisc.TestID.
Iftheinformationisincorrect:
SelectCORRECTMISC.TESTIDusingthekeys.
Pressthekey.
MainMenu
Run Test
NOTE
:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
NOTE
:IftheBarCodeScannerisbeingused,simplypointthescanneratthebarcode,pressthebutton
onthescannerhandleandwaitforthebeep.TheIDwillappearonthedisplay.