- 31 -
Once the user name is added, enter the following:
User code Max 6 characters,
User password Max 8 numbers.
Next select:
Access level (USER, ADV, ADMIN)
Language
In order to remove a user you should follow this procedure:
Enter the user database
Select the user that is to be removed from the list
Press button
The software shows <DELETE?> message in the bottom line
Confirm by pressing button
When confirmed, the software removes the selected user from the list
9.2. PRODUCTS
PRODUCTS1000 different products.
The following data can be inserted for each product:
NAME (30 characters),
CODE (6 characters),
EAN (16 characters),
MASS (with the accuracy of a reading unit),
TARE (mass of the packaging relating to a particular product with the accuracy of a reading
unit),
MIN (low limit for the <CHECKWEIGHING> mode, to be inserted with the accuracy of a reading
unit),
MAX (high limit for the <CHECKWEIGHING> mode, to be inserted with the accuracy of a
reading unit),
TOLERANCE (tolerance limits to [±] for <DOSING> mode, entered as a % of the target mass).
In order to add a product, enter products database and add name of the product along with
respective data (follow the procedure as in the section above).

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals