EasyManuals Logo

RADWAG AS R2 PLUS User Manual

RADWAG AS R2 PLUS
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
- 76 -
value NO’ – cooperation with other programs (mass value
transferred in a form of text),
DELAY if the transferred information is incomplete, i.e. part of it gets lost (due to auto-
entering or auto-formatting of data by computer program) during cooperation with Excel
type program, set the parameter value to <2>. If the transferred information is still
incomplete, the parameter value must be higher. The value ranges between 0 (quick data
transfer) and 9 (slow data transfer around 10 characters per second).
In case of Free Link printout (keyboard emulator) all languages without special characters
are operated (‘QWERTY’ keyboard layout), and two languages that feature special
characters:
Polish for ‘Polish (Programmers) keyboard’. To enable support of this language go to
printer settings on your balance and select Windows-1250 code page (Central
European, Latin-2). The following Polish diacritical signs are added: ą, ć, ę, ł, ń, ó, ś, ż,
ź.
German for ‘German (Germany) keyboard’. To enable support of this language go to
printer settings on your balance and select Windows -1252 code page (Western
European, Latin-1). The following German diacritical signs are added: ä, ö, ü, ß. The
keyboard layout changes from ‘QWERTY’ to ‘QWERTZ’. The characteristic feature of
such keyboard is ‘ß’ sign located on the right side of the ‘zero’.
Caution!
Free Link printout option allows printing only signs that are available on a keyboard.
Entering capital letters and special signs, by means of key combination where "Shift"
key is pressed, is also possible.
In case of Polish and German, basic special characters (written using key combination
of a given letter and right ‘Alt’ button) have been added.
Free Link printout option does not support printing of remaining special signs.
WIFIWIFI port
13.1.2. Continuous Transmission
Continuous transmission parameter enables turning on/off continuous transmission of
a measurement result.
Available options:
Continuous transmission in a calibration unit <CAL. UNIT>, independently from currently
chosen weight unit, measurements are transmitted in unit set on a main measuring device
(calibration unit)
Continuous transmission in a current unit <CURRENT UNIT> measurements are transmitted in
a current unit; the unit changes dynamically whenever the user presses Units button in order to
modify the unit on a display.
Continuous transmission off <NONE>
Procedure:

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals