11.2. WEIGHING OPERATION
11.2.1. Good Weighing Practice
In order to ensure long lasting use of a balance plus correct and reliable measurement of weighed
loads, follow below procedures:
Start the balance with no load on the weighing pan (permissible value of load on the weighing
pan on balance start is ±10% of its maximum capacity).
Load the weighing pan steadily avoiding shocks:
YES
NO
Place weighed loads centrally on the weighing pan:
YES
NO
Avoid side loading, in particular side shocks:
NO
NO
The balance requires adjusting before weighing process start or in case of drastic change of ambient
conditions at a workstation.
Before the start of weighing procedure, it is recommended to load the weighing pan a few times
with mass close to balance max capacity,
Check if unloaded balance indicates “precise zero” and whether measurement is stable
, if not, press /Delete button,
Press UNITS button, to set a measuring unit:
Place weighed object on the weighing pan and read result only on measurement result
stabilization,
Mass indication of a load placed on the weighing pan can be tared multiple number of times by
pressing T
/Insert (pay CAUTION not to exceed maximal capacity of a balance by applying
multiple tare function).
The balance shall stay plugged to the mains in between measurement series. It is recommended to
switch off the balance display by pressing ON/OFF button. On repeated ON/OFF button pressing, the
balance is ready for operation and carrying out the following measurements.

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals