- 85 -
XX _ v CR LF
command understood but min threshold is exceeded
XX _ OK CR LF
command carried out
ES_CR LF
command not recognised
XX _ E CR LF
time limit exceeded while waiting for stable measurement result
(time limit is balance characteristic parameter)
Where:
XX - name of a sent command
_ - spaces
Commands Overview
Zero balance
Format: Z CR LF
Response options:
Z_A CR LF
- command understood and in progress
Z_D CR LF
- command carried out
Z_A CR LF
- command understood and in progress
Z_^ CR LF
- command understood but zeroing range is exceeded
Z_A CR LF
- command understood and in progress
Z_E CR LF
- time limit exceeded while waiting for stable measurement result
Z_I CR LF
- command understood but not accessible at this moment
Tare balance
Format: T CR LF
Response options:
T_A CR LF
- command understood and in progress
T_D CR LF
- command carried out
T_A CR LF
- command understood and in progress
T_v CR LF
- command understood but taring range exceeded
T_A CR LF
- command understood and in progress
T_E CR LF
- time limit exceeded while waiting for stable measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format: OT CR LF
Response: OT_TARE CR LF - command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
O T space tare space unit space CR LF
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
CAUTION!
Tare value is always given in calibration unit.
Set tare
Format: UT_TARE CR LF, where TAREtare value
Response options:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION!
Use dot in tare format as decimal point.
Send stable measurement result in a basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals