Page #86 background imageLoading...
Page #86 background image
- 86 -
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space
stability
marker
space character mass space unit CR LF
An example:
S CR LF
- command send form a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
- command carried out, response: mass value in a basic
measuring unit.
where: _ - space
Immediately send measurement result in a basic measuring unit
Format: SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a basic measuring unit
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
An example:
S I CR LF
command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command carried out, immediate response: mass
value in a basic measuring unit
where: _ - space
Send stable measurement result in a current measuring unit
Format: SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
SU_E CR LF
- time limit exceeded while waiting for a stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U space
stability
marker
space character mass space unit CR LF
An example:
S U CR LF
- command send form a computer
S U _ A CR LF
- command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command carried out, response: mass value in a
current measuring unit
where: _ - space
Immediately send measurement result in a current measuring unit
Format: SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
An example:
S U I CR LF
command sent from a computer

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals