S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate response: mass value
in a current measuring unit
Switch on continuous transmission in a basic measuring unit
- command understood but not accessible at this moment
- command understood and in progress
- response: mass value in a basic measuring unit
S I space
space character mass space unit CR LF
Switch off continuous transmission in a basic measuring unit
- command understood but not accessible at this moment
- command understood and carried out
Switch on continuous transmission in a current measuring unit
- command understood but not accessible at this moment
- command understood and in progress
- response: mass value in a current measuring unit
S U I
space character mass space unit CR LF
Switch off continuous transmission in a current measuring unit
- command understood but not accessible at this moment
- command understood and carried out
Set min checkweighing threshold
Format: DH_XXXXX CR LF, where: _ - space, XXXXX – mass format
- command not recognised (mass format incorrect)
Set max checkweighing threshold
Format: UH_XXXXX CR LF, where: _ - space, XXXXX - mass format
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Response: DH_MASA CR LF - command carried out
1 2 3 4-12 13 14 15 16 17 18 19
D H space mass space unit space CR LF
- 9 characters with right justification
- 3 characters with left justification
Give value of max checkweighing threshold