- 88 -
Response: UH_MASA CR LF - command carried out
1 2 3 4-12 13 14 15 16 17 18 19
U H space mass space unit space CR LF
- 9 characters with right justification
- 3 characters with left justification
Set mass value of a single item (only for PARTS COUNTING)
Format: SM_XXXXX CR LF, where: _ - space, XXXXX – mass format
SM_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
PARTS COUNTING)
- command not recognised (mass format incorrect)
Set target mass value (e.g. for DOSING)
Format: TV_XXXXX CR LF, where: _ - space, XXXXX - mass format
TV_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
DOSING)
- command not recognised (mass format incorrect)
Set reference mass value (e.g. for PERCENT WEIGHING)
Format:
, where:
space
– mass format
Response options:
RM_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
PERCENT WEIGHING)
- command not recognised (mass format incorrect)
Value release
- command understood and in progress
Command’s function is similar to function of PRINT button located on an overlay,
Internal adjustment
- command understood and in progress
- command understood and in progress
- time limit exceeded while waiting for a stable measurement result
- command understood but not accessible at this moment
Disable automatic internal adjustment of the balance
- command understood but not accessible at this moment
- operation disabled, e.g. for verified balance
Operation is disabled for verified balances.
For non-
verified balances the command inhibits internal calibration until it is enabled via IC0
command or until the balance is turned off. The command does not modify settings specifying
calibration start.
Enable automatic internal adjustment of the balance