- 90 -
OK <CR><LF>
command carried out
An example 2:
Command:
OMI <CR><LF>
return accessible working modes
Response:
OMI <CR><LF>
2 <CR><LF>
4 <CR><LF>
12 <CR><LF>
accessible working modes are given in
return:
mode number
OK <CR><LF>
command carried out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n <CR><LF>
Response options:
OMS_OK <CR><LF>
command carried out
OMS_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
OMS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_13<CR><LF>
set Statistics mode
Response:
OMS_OK<CR><LF>
Statistics mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG <CR><LF>
Response options:
OMG_n_OK <CR><LF>
command carried out, response: current working mode
OMG_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command.
An example:
Command:
OMG<CR><LF>
return current working mode
Response:
OMG_13_OK<CR><LF>
balance operates in Statistics working mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current
working mode.
Format: UI <CR><LF>
Response options:
UI_”x
1
,x
2
, … x
n
”_OK<CR><LF>
command carried out, returns accessible units
UI_I <CR><LF>
command understood but not accessible at this
moment
x unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI <CR><LF>
return available units
Response:
UI_”g, mg, ct”_OK<CR><LF>
response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x <CR><LF>
Response options:
US_ x_OK <CR><LF
command carried out, response: current unit
US_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
US_I <CR><LF>
command understood but not accessible at this moment
x parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „ ” button

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals