- 91 -
pressing).
An example:
Command:
US_mg<CR><LF>
set „mg” unit
Response:
US_mg_OK<CR><LF>
„mg” set as a current unit
Give current unit
Command overview: command returns current unit.
Format: UG <CR><LF>
Response options:
UG_x_OK<CR><LF>
command carried out, response: current unit
UG_I <CR><LF>
command understood but not accessible at this moment
x parameter, unit symbol
An example:
Command:
UG<CR><LF>
return current unit
Response:
UG_ct_OK<CR><LF>
currently set unit is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS <CR><LF>
Response options:
BP_OK <CR><LF>
command carried out, BEEP sound signal activated
BP_E” <CR><LF>
no parameter or incorrect format
BP_I <CR><LF>
command understood but not accessible at this moment
TIME parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 - 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350<CR><LF>
activate BEEP for 350 ms
Response:
BP_OK<CR><LF>
BEEP activated
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation
the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
Send all implemented commands
Format: PC CR LF
Command:
PC CR LF
- send all implemented commands
Response:
PC_A_”Z,T,S,SI…”
- command carried out, the terminal displays all
implemented commands.
Give balance type
Format: BN <CR><LF>
Response options:
BN_A_”x” <CR><LF>
- command understood, response: balance type
BN_I <CR><LF>
- command understood but not accessible at this moment
x series of types for a particular balance (in between inverted commas), with general balance type
in front
An example:
Command:
BN <CR><LF>
return balance type
Response:
BN_A_”AS”
balance type: ”AS R”
Give max capacity
Format: FS <CR><LF>
Response options:
FS_A_”x” <CR><LF>
- command understood, response: Max capacity
FS_I <CR><LF>
- command understood but not accessible at this moment
xMax value of reading units (in between inverted commas)
An example:
Command:
FS <CR><LF>
return Max capacity
Response:
FS_A_”220.0000”
Max capacity: ”220 g”
Give program version

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals