Format: RV <CR><LF>
Response options:
RV_A_”x” <CR><LF>
- command understood, response: program version
RV_I <CR><LF>
- command understood but not accessible at this moment
x – program version (in between inverted commas)
An example:
Command:
RV <CR><LF>
return program version
Response:
RV_A_” 1.1.1”
program version: ”1.1.1”
Set AUTOZERO function
Format: A_n <CR><LF>
Response options:
A_OK <CR><LF>
command carried out
A_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
A_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining autozero settings
n → 0 autozero off
1 autozero on
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
A_1<CR><LF>
turn autozero function on
Response:
A_OK<CR><LF>
autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format: EV_n <CR><LF>
Response options:
EV_OK <CR><LF>
command carried out
EV_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
EV_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining ambient conditions state
n → 0 unstable ambient conditions
1 stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1<CR><LF>
set value ‘stable’ for ambient
conditions option
Response:
EV_OK<CR><LF>
ambient conditions option set to
value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to value
<UNSTABLE>.
Give currently set ambient conditions
Command overview:
On receipt of a command, the terminal sends information on currently set ambient conditions.
Format:
EVG <CR><LF>
Response options:
EVG_x_OK<CR><LF>
- command carried out, returns set ambient conditions
EVG_I <CR><LF>
- command understood but not accessible at this moment
x - parameter, ambient conditions symbols
Example:
Command:
EVG<CR><LF>
- give currently set ambient conditions

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals