- 93 -
Response:
EVG_0_OK<CR><LF>
- currently set ambient conditions unstable
Set filter
Format: FIS_n <CR><LF>
Response options:
FIS_OK <CR><LF>
command carried out
FIS_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
FIS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining filter number
n → 1 very fast
2 fast
3 average
4 slow
5 very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current wor
king mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3<CR><LF>
set average filter
Response:
FIS_OK<CR><LF>
average filter set
Give current filter
Command overview:
On receipt of a command, the terminal sends information on currently set filter.
Format:
FIG <CR><LF>
Response options:
FIG_x_OK<CR><LF>
- command carried out, returns set filter
FIG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, filter symbols
Example:
Command:
FIG<CR><LF>
- give current
filter
Response:
FIG_2_OK<CR><LF>
- currently set
average filter
Set Value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
command carried out
ARS_E <CR><LF>
error in-course of command execution, no parameter or
incorrect format
ARS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining value release options
n → 1 fast
2 fast+reliable
3 reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release settings
are assigned to the working mode
An example:
Command:
ARS_2<CR><LF>
set value release parameter to
fast+reliable option
Response:
ARS_OK<CR><LF>
fast+reliable option set

Table of Contents

Other manuals for RADWAG AS R2 PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 PLUS and is the answer not in the manual?

RADWAG AS R2 PLUS Specifications

General IconGeneral
BrandRADWAG
ModelAS R2 PLUS
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals