VARIABLE 1 – function initiates the selection and editing of variable 1
VARIABLE 2 – function initiates the selection and editing of variable 2
The way of declaring the function is described in the section relating to
the weighing mode settings <F button shortcuts>.
17.5.1. Setup for reference mass determined by weighing
Procedure:
• Press one of the F buttons, to which the function <DETERMINE
SAMPLE> is assigned and the editing window will be displayed
<PLACE 100%>
• Place the standard (as 100%) on the pan, and when the result is stable
(
symbol is displayed) confirm the mass by pressing
• The balance program automatically enters the load value to be
measured as a standard and enters the mode <DEVIATIONS>
displaying the value of 100.000%, and the bottom line displays
standard mass value (if the option for function <INFORMATION> is
selected).