20.2. Format of data setting
The measurement result can be sent from the balance to an external device by
pressing
button on the balance or through sending a command from the
computer.
20.3. Format of data sent on pressing of PRINT button
CAUTION
Unstable measurement print-out is not available for a verified balance.
Printout format
Mass
space
unit
CR
LF
stability marker - [space] – when stable
- [?] – when unstable
- [^] - if there is an error exceeding the maximal range +
- [v] - if there is an error exceeding the maximal range -
character - [space] – for positive or [-] negative values
mass - 9 characters – aligned to the right
unit - 3 characters – aligned to the left.
20.3.1. Format of data sent as response for commands generated
from a computer
After entering a command balance responds to the following first:
XX_A CR LF - command understood and in progress
XX_I CR LF - command understood but not accessible at this
moment
XX_^ CR LF - command understood but maximum range is exceeded
XX_v CR LF - command understood but minimum range is exceeded
XX_E CR LF - an error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result),
time limit is balance’s characteristic parameter
XX - name of the command