13.3.2. External adjustment
The external adjustment in balances NPS series should be carried out with
external mass standard / weigh class F
1
Start of external adjustment process,
Balance will display a command ordering unloading of weighing pan
<REMOVE MASS> (weighing pan must be empty). When weighing pan is
unloaded, press
button.
The balance determines mass of empty pan and message <CALIBRATION>
is displayed in the bottom line. It is followed by message message <PLACE
MASS> in the bottom bar, and in the main window the mass value to be
placed on the weighing pan e.g. 200.000g (depending on the type of
balance). Place an external adjustment weight with mass determined on
balance’s display and press
button. The balance determines the mass
and message <CALIBRATION> is displayed in the bottom line. On
completing adjustment process the balance returns to submenu P1.2
INT.CALIB.
If a balance is verified, it is not possible for a user to carry out
external adjustment process.
13.3.3. User adjustment
The external adjustment in balances NPS series should be carried out with
external mass standard / weigh class F
1
Start of external adjustment process,

Table of Contents