Automatic operation proceeds as follows:
Press button to zero the balance (marker of stable
measurement
and zero marker are shown on display)
Place the load, the balance sends the first stable measurement to
the printer port,
Remove the load from the balance,
Next measurement is possible when the reading on the display is
lower than the set value of < AUTO THRES. > parameter (next
measurement does not require zero value).
This procedure applies to changing settings:
For automatic working mode adjust the threshold value.
< AUTO THRES. > parameter is connected with the following functions:
- automatic tare,
- automatic operation,
For automatic operation function the measurement will not be send from the
computer to the printer as long as the mass measurement stays withing the set
value range < P 2.2.3.2 PROG THRES>.