EasyManuals Logo

RADWAG PS Series User Manual

RADWAG PS Series
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
16
7. START-UP
When plugged to mains, the balance displays name and number of software, and
afterwards proceeds to weighing mode.
7.1 Balance temperature stabilization period
Before start of measuring processes, it is necessary to wait until the balance is
thermally stabilized. It is a period of so called balance self-heating. In case of
balances PS series self-heating period takes approximately 30 minutes. The
specified time interval refers to balances that have been stored in room temperature
before plugging to mains.
For balances that were stored in much lower temperatures before plugging to mains
(e.g. during winter period) thermal stabilization should last approximately 4 hours.
Correct operation of a balance is possible within temperature range specified in the
technical parameters, see point 1. If the balance is placed in area which ambient
temperature is out of temperature range specified in the technical data, then
balance’s display may signal such situation by lighting an appropriate marker (a
black triangle).
In case the balance displays the marker, increase or decrease ambient
temperature, depending on present temperature value. It is recommended that
ambient temperature changes at balance’s place of use are very small (slow).

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
CategoryScales
BrandRADWAG
SeriesPS
CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
DisplayLCD
HousingPlastic
RepeatabilityVaries by model
LinearityVaries by model
CalibrationInternal or external
InterfaceRS232
Power Supply100 - 240V AC
Operating Temperature+15 ÷ +30 °C
Weighing Unitg, lb, oz

Related product manuals