12.5. Adjustment report printout
On completion of any type of adjustment process, the balance enables
preparing a report from adjustment process. The report can be printed on a
connected printer and sent to a computer and saved in a form of file for
records. Printout process is described in details in point 9.2 of this user manual.
The user can select between two options.
no – report is not printed
YES – report is printed
Remember, that if the parameter is set for YES, then a report is generated and
sent automatically.
Content of report from adjustment process depends on settings of GLP
parameters. Any option in the GLP submenu which attribute is YES is included
in a report from adjustment process
P2 GLP (Good Laboratory Practice)
P2.1 uSr |
P2.2 PrJ |
P2.3 Ptin | YES
P2.4 PdAt | YES
P2.5 PuS | YES
P2.6 PPrJ | YES
P2.7 PId | YES
P2.8 PFrn | YES
Apart from information set in menu group the report contains:
• Mass of adjustment weight stored in balance’s memory from last
carried out adjustment;
• Mass of adjustment weight determined in current adjustment process;
• Adjustment deviation, i.e. difference between the two mass records.