P1.1 iCAL Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically, and
no operator assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with relativaly small mass (up to 4 % of balance max capacity) than
adjustment process will be carried out automatically, but measuring results may
differ by the amount that was present on balance weighing pan.
or CAL function key
P1.2 ECAL External adjustment (with an external weight)
adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
balances with internal adjustment balances with external adjustment
P1.3 tCAL Adjustment test
mass comparison of internal adjustment weight with its value saved in balance
memory.
balances with internal adjustment balances with external adjustment

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
CategoryScales
BrandRADWAG
SeriesPS
CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
DisplayLCD
HousingPlastic
RepeatabilityVaries by model
LinearityVaries by model
CalibrationInternal or external
InterfaceRS232
Power Supply100 - 240V AC
Operating Temperature+15 ÷ +30 °C
Weighing Unitg, lb, oz

Related product manuals