Some general conditions apply to event-based actions:
Rules are defined globally. However, you must activate each rule for each applica-
tion individually.
Only the currently active measurement channel processes rules.
The R&S FSV/A in general, or any applications, do not perform a validity check on
the rules.
When you save and then recall the instrument settings, any rule definitions are also
stored and loaded, including their state.
After an instrument preset or restart, the rules are maintained, but disabled. The
trigger count for rules is cleared.
Managing Rules.................................................................................................... 581
Defining Rules.......................................................................................................582
Debugging Rules with the EBA-Journal................................................................588
Reference: Overview of Available Result-Events..................................................589
12.3.1 Managing Rules
Access: toolbar
The event-based actions manager allows you to configure actions to be performed
when specific events occur in the measurement.
Rule State................................................................................................................... 582
Rule.............................................................................................................................582
Events......................................................................................................................... 582
Actions........................................................................................................................ 582
Triggered.....................................................................................................................582
New Rule.....................................................................................................................582
Edit Rule......................................................................................................................582
Copy Rule................................................................................................................... 582
Delete Rule................................................................................................................. 582
Show Journal.............................................................................................................. 582
Event-based Actions

Table of Contents

Other manuals for R&S FSV3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3030 and is the answer not in the manual?

R&S FSV3030 Specifications

General IconGeneral
BrandR&S
ModelFSV3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals