EasyManuals Logo

R&S FSV3030 User Manual

R&S FSV3030
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #707 background imageLoading...
Page #707 background image
Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
707User Manual 1178.8520.02 ─ 01
"VXI-11 Dis-
covery"
If enabled, connected devices are detected automatically using the
VXI-11 protocol (see "VXI-11 Protocol" on page 647)
Remote command:
SYSTem:LXI:INFO? on page 1125
LXI Password
Password for LAN configuration. The default password is LxiWebIfc.
Remote command:
SYSTem:LXI:PASSword on page 1126
LXI Manufacturer Description
Instrument description of the R&S FSV/A
Remote command:
SYSTem:LXI:MDEScription on page 1126
LAN Reset
Resets the LAN configuration to its default settings (LCI function).
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter Value
TCP/IP Mode DHCP + Auto IP Address
Dynamic DNS Enabled
ICMP Ping Enabled
Password for LAN configuration LxiWebIfc
Remote command:
SYSTem:LXI:LANReset on page 1126
13.5.5 Remote Errors
Access: [SETUP] > "Network + Remote" > "Remote Errors " tab
The error messages generated by the R&S FSV/A during remote operation are dis-
played here.
The messages are displayed in the order of their occurrence; the most recent mes-
sages are placed at the top of the list.
Network and Remote Control Settings

Table of Contents

Other manuals for R&S FSV3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3030 and is the answer not in the manual?

R&S FSV3030 Specifications

General IconGeneral
Frequency Range10 Hz to 30 GHz
Input Impedance50 Ω
Input VSWR< 1.5:1
Input Power Level+30 dBm
Display1280 x 800 pixels
Power Supply100 V to 240 V, 50 Hz to 60 Hz
Resolution Bandwidth1 Hz to 10 MHz (1 Hz to 50 MHz with option)
Displayed Average Noise Level (DANL)-165 dBm (1 Hz RBW)

Related product manuals