Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1085User Manual 1178.8520.02 ─ 01
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
on page 1084and
SYSTem:COMMunicate:PRINter:ENUMerate[:NEXT]?
on page 1084
to query all available printers.
*RST: NONE
Manual operation: See " Printer Name " on page 552
14.9.5 Storing Measurement Results
The following commands can be used to store the results of a measurement.
Useful commands for storing results described elsewhere:
FORMat[:DATA] on page 975
MMEMory:STORe<n>:TRACe on page 979
FORMat:DEXPort:TRACes on page 978
Remote commands exclusive to storing results:
FORMat:DEXPort:HEADer............................................................................................1085
MMEMory:STORe<n>:LIST...........................................................................................1085
MMEMory:STORe<n>:PEAK.........................................................................................1086
MMEMory:STORe<n>:SGRam......................................................................................1086
MMEMory:STORe<n>:SPECtrogram............................................................................. 1086
MMEMory:STORe<n>:SPURious.................................................................................. 1087
FORMat:DEXPort:HEADer <State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
See Chapter 9.6.6, "Reference: ASCII File Export Format", on page 525 for details.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Manual operation: See " Include Instrument & Measurement Settings " on page 521
MMEMory:STORe<n>:LIST <FileName>
This command exports the SEM and spurious emission list evaluation to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
Managing Settings and Results

Table of Contents

Other manuals for R&S FSV3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3030 and is the answer not in the manual?

R&S FSV3030 Specifications

General IconGeneral
BrandR&S
ModelFSV3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals