Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
921User Manual 1178.8520.02 ─ 01
See "Adjusting settings automatically during triggered measurements" on page 409.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Manual operation: See " Adjusting the Center Frequency Automatically ( Auto Fre-
quency )" on page 410
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S FSV/A or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Manual operation: See " Setting the Reference Level Automatically ( Auto Level )"
on page 371
14.7.6 Configuring the Data Input and Output
The following commands are required to configure data input and output.
RF Input................................................................................................................ 921
Setting up Probes..................................................................................................925
External Generator Control................................................................................... 930
Working with Power Sensors................................................................................ 941
Configuring the Outputs........................................................................................ 951
14.7.6.1 RF Input
CALibration:PADJust[:STATe].......................................................................................... 922
INPut<ip>:ATTenuation:PROTection:RESet.......................................................................922
INPut<ip>:COUPling.......................................................................................................922
INPut<ip>:DPATh...........................................................................................................923
INPut<ip>:FILTer:YIG[:STATe]..........................................................................................923
Setting Basic Measurement Parameters

Table of Contents

Other manuals for R&S FSV3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3030 and is the answer not in the manual?

R&S FSV3030 Specifications

General IconGeneral
BrandR&S
ModelFSV3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals